Resonance Frequency Analysis of Dental Implants Placed Into the Sinus Lifts: at Stage1 and Stage 2 Surgery
- 1 Department of Dentistry, Charles University in Prague, Faculty of Medicine in Hradec Králové Head of the department doc. MUDr. V. Hubková, CSc.
- 2 Department of Biophysics, Charles University in Prague, Faculty of Medicine in Hradec Králové Head of the department prof. MUDr. P. Stránský, CSc.
Measurement of primary stability of dental implants using resonance frequency analysis is found to be a reliable and evidence based method. It is believed that a higher value indicates a better stability and so a better prognosis for the implant. This kind of relation can also be expected from implants inserted in sinus lifts. This study proves that the values with a lower amplitude by time reach a higher value and those with higher initial resonance frequency value tend to decrease to somewhat lower values. That is, the secondary stability of the implant does not depend completely on primary stability value.
Keywords: sinus lift; resonance frequency analysis; ISQ values; implant stability
Published: December 1, 2006 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (CC BY-NC 4.0), which permits non-comercial use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.